How to Play

Press A to turn on the aerosol.

The client's SOP states the minimum upstream concentration should be between 12 and 20 µg/L.

Adjust the upstream concentration with the slider or by holding + or -.

To begin scanning the filter press Space. Overlapping strokes will be displayed as a darker green. Be sure not to go too fast!!

If you give up on finding the leak(s) press L to expose them all.

Key Features

  • Practice standard test methods including scan speed and overlap.
  • Variable Leak Geometry: Detect pinhole leaks, gasket seal failures, and frame bypasses.
  • Certification Workflows: Complete upstream challenge verification and 100% concentration baselining.

Training Objectives

Develop muscle memory for proper probe handling (approx. 2 in/s) without the risk of damaging expensive PTFE or Glass Fiber media. Master the coordination required for PAO/DOP aerosol injection and simultaneous scanning in a safe, virtual environment.